Nebulizer
Systems Frequently Asked Questions
How
can I clean the main unit of the Nebulizer System?
To clean the casing of the main unit moisten a soft cloth with water or
a mild detergent. Wipe the casing and immediately dry using a soft clean
cloth. Do not rinse or immerse the Main Unit in any liquid or clean the
Main Unit using abrasive cleaners or any type of chemical. (Back
to Top)
How much medication will the medication cup hold
in the Omron MicroAir Electronic Nebulizer System?
The medication cup on the MicroAir NE-U03V holds 5 ml. The medication
cup on the MicroAir NE-U22V holds 7 ml. (Back to Top)
How often should the filter be replaced on the
CompAir or CompAir XL Compressor Nebulizer System?
Change the filter every 30 days or when the filter turns grey. The CompAir
and the CompAir XL use replacement filters model number 9930. Filters
are sold 5 in a package. (Back to Top)
How often should the filter be replaced on the Omron CompAir
Elite Compressor Nebulizer System?
The filter for the NE-C21 CompAir Elite must be replaced after one month
of use, if the filter has turned grey, or the device has been in storage
for three months or longer. Replacement filters are sold 5 in a package,
model number 9935. (Back to Top)
How often should the nebulizer kit be changed
on the Omron CompAir Elite Compressor Nebulizer System?
The nebulizer kit for the NE-C21 CompAir Elite should be replaced after
six months of use, if the nebulizer kit does not operate properly, or
after the device has been in storage for a while. (Back to
Top)
Is a European Adapter available for the Omron
MicroAir NE-U03V Nebulizer System?
You can purchase the AC Adapter (220v, 50Hz) as an optional accessory.
The part number is U1002. (Back to Top)
Should the Omron MicroAir Electronic Nebulizer
System be used to take treatments with the continuous or intermittent
flow?
The MicroAir Electronic Nebulizer System is designed for both types of
operation to accommodate your needs. Follow the instructions of your healthcare
provider regarding your treatment plan. (Back to Top)
The CompAir Elite is on, but nebulizes weakly. What would
cause this?
Make sure the nebulizer kit is seated correctly in the Main Unit of the
NE-C21 and that the air tubing connection is free of obstruction. The
nebulizer should be in an upright position during use. Do not tilt beyond
30º. Check whether the nebulizer is empty. You may need to clean
or replace the nebulizer kit. Check the air filter to see if it's dirty
or obstructed. (Back to Top)
What material is the pediatric mask used with
the Omron CompAir Elite Compressor Nebulizer System made of?
The pediatric mask, model number 9971, used with the Omron NE-C21 CompAir
Elite is made of poly vinyl chloride. (Back to Top)
Why does the Omron CompAir Elite shut down when
the indicator light turns red?
The NE-C21 CompAir Elite has an internal thermal switch to protect the
unit from overheating. The thermal protector has been activated because
the device has been operating beyond the operating conditions indicated
in your instruction manual, or the device has been operating near heat
sources, or in hot environments. Turn the unit off for at least 20 minutes
to allow the unit to cool down. Wait until the red light turns off and
turn the unit on again. (Back